Specifications
| Length | Code 0 (A) | Code 5 (A) | Code 6 (A) | Code 7 (A) |
| 5" | 148.0 mm (5.8") | – | 150.5 mm (5.9") | – |
| 10" | 273.0 mm (10.7") | 314.0 mm (12.4") | 276.0 mm (10.9") | 316.5 mm (12.5") |
| 20" | 517.5 mm (20.4") | 558.5 mm (22.0") | 520.0 mm (20.5") | 561.0 mm (22.1") |
| 30" | 762.0 mm (30.0") | 803.0 mm (31.6") | 764.5 mm (30.1") | 805.5 mm (31.7") |
| 40" | 1006.5 mm (39.6") | 1047.5 mm (41.2") | 1009.0 mm (39.7") | 1050.0 mm (41.3") |
| Materials of construction | Membrane | Hydrophilic PES |
| Support | Polypropylene | |
| Core and Cage | Polypropylene | |
| End Caps | Polypropylene with stainless steel (SS) lining | |
| O-rings/Gasket | Silicone/EPDM | |
| Sealing technology | Thermal bonding, no adhesives | |
| Dimensions |
Diameter | Ø70 mm |
| Length | 5”, 10”, 20”, 30”, 40” | |
| Filtration area | ≥6.4 ft2 per 10” cartridge | |
| Retention rating | 0.1, 0.22, 0.45, 0.22+0.1, 0.22+0.22, 0.45+0.22 μm (Other retention rating on request) | |
| Maximum differential pressure | Forward: | 4.2 bar @ 23°C (73°F), 1.0 bar @ 85°C (185°F) |
| Integrity test-water bubble point at 23°C | 0.1 μm | ≥3800 mbar |
| 0.22 μm | ≥3400 mbar | |
| 0.45 μm | ≥2300 mbar | |
| Bacterial retention (for 0.1 μm & 0.22 μm only) | Passed the bacterial challenge testing using Brevundimonas diminuta (ATCC19146) at a minimum challenge concentration of 1 × 107 CFU/cm2 | |
| Multiple sterilization cycles |
May be steam-in-place sterilization for maximum 20 cycles at 127°C (261°F), 30 min, ΔP≤300mbar | |
| May be autoclaved for maximum 20 cycles at 123°C (253°F), 30 min | ||
| May be sanitized by hot water, maximum 30 cycles at 85°C (185°F) | ||
| Food contact material | Meets the requirements of (EC) No 1935/2004 | |
| Indirect food additive | Meets the FDA Indirect Food Additive requirements cited in 21 CFR 177-182 | |
| Quality | Manufactured in accordance with ISO 9001 and ISO 13485 certified quality management system | |